One step calibration of industrial hyperspectral cameras

Martin L. Henriksen, William N. Pedersen, Pernille Klarskov, Mogens Hinge*

*Corresponding author for this work

Research output: Contribution to journal/Conference contribution in journal/Contribution to newspaperJournal articleResearchpeer-review

68 Downloads (Pure)

Abstract

Hyperspectral imaging is gaining interest as an inline reflective analytical technique. The raw data is often applied for classification or process adjustments. Proper calibration can eliminate the background, noise, and artefacts. Further, it enables determination of area fractions and, via spectral information, chemical analysis. This paper demonstrates a simple, robust, and easy-to-operate calibration device for inline calibration of hyperspectral cameras in the range from 450 ​nm to 1840 ​nm. The design is flexible, made from easy sourced materials, and can be adapted to most industrial setups. Calibration of intensity, spatial and spectral information is combined in one device and is done in a single pass. Device alignment and subsequent image analysis provides a fast and easy calibration routine for hyperspectral cameras.

Original languageEnglish
Article number104609
JournalChemometrics and Intelligent Laboratory Systems
Volume227
Number of pages7
ISSN0169-7439
DOIs
Publication statusPublished - Aug 2022

Keywords

  • Calibration method
  • Hyperspectral imaging
  • Mechanical design
  • Signal processing

Fingerprint