Influence of metal artefact reduction on the diagnosis of contact between implant and mandibular canal in cone beam computed tomography: An ex-vivo study

Bruna Neves de Freitas, Raphael Jurca Gonçalves da Motta, Ruben Pauwels, Christiano Oliveira-Santos, Camila Tirapelli*

*Corresponding author for this work

Research output: Contribution to journal/Conference contribution in journal/Contribution to newspaperJournal articleResearchpeer-review

5 Citations (Scopus)

Abstract

Objective: To evaluate the influence of metal artefact reduction (MAR) in the diagnosis of dental implant contact with the mandibular canal (MC) using cone beam computed tomography (CBCT). Methods: Dental implants were installed with surgical guides in the posterior hemiarches of 10 dry human mandibles: 0.5 mm above to the MC cortex (G1/n = 8) and 0.5 mm inside the MC (G2/n = 10). The experimental set-up was scanned with two CBCT equipment using 85 kV and 90 kV, MAR ON or OFF, and different tube currents (4 mA, 8 mA and 10 mA). Two dentomaxillofacial radiologists (DMFRs) and two dentists (DDS) scored the relation between the dental implant and MC. Descriptive statistics were used to observe the absolute frequency of scores. Sensitivity, specificity and accuracy were calculated considering the known relation between the dental implant and the MC interior. McNemar's test (α =.05) was applied to compare the diagnostic efficacy of MAR ON versus MAR OFF. Results: Overall specificity was higher than sensitivity for both DDS and DMFR (97% vs. 50% and 92.0% vs. 78.0% respectively). There was a significant effect of MAR (p =.031) for DMFR in the case of contact between the dental implant with the MC interior, in which sensitivity decreased with MAR activation from 90% to 40%. DMFR observers showed a better diagnostic performance compared with the DDS observers (accuracy of 84.0% and 71.0%, respectively). Conclusions: Due to the limited efficacy of MAR, it should not be used when conducting CBCT scans for the evaluation of contact between the implant and the mandibular canal.

Original languageEnglish
JournalClinical Oral Implants Research
Volume34
Issue7
Pages (from-to)741-750
Number of pages10
ISSN0905-7161
DOIs
Publication statusPublished - Jul 2023

Keywords

  • cone beam computed tomography
  • dental implants
  • diagnostic imaging
  • mandibular canal

Fingerprint

Dive into the research topics of 'Influence of metal artefact reduction on the diagnosis of contact between implant and mandibular canal in cone beam computed tomography: An ex-vivo study'. Together they form a unique fingerprint.
  • Honorable Mention - Oral Presentation

    de Freitas, B. N. (Recipient), da Motta, R. J. G. (Recipient), Pauwels, R. (Recipient), Oliveira-Santos, C. (Recipient) & Tirapelli, C. (Recipient), 5 Sept 2023

    Prize: Prizes, scholarships, distinctions

Cite this