Extracellular hydrogen peroxide measurements using a flow injection system in combination with microdialysis probes – Potential and challenges

Research output: Contribution to journal/Conference contribution in journal/Contribution to newspaperJournal articleResearchpeer-review

  • Maria Moßhammer, Københavns Universitet
  • ,
  • Verena Schrameyer, Københavns Universitet
  • ,
  • Peter Jensen, Rigshospitalet, Københavns Universitet
  • ,
  • Klaus Koren
  • Michael Kühl

There is a strong need for techniques that can quantify the important reactive oxygen species hydrogen peroxide (H2O2) in complex media and in vivo. We combined chemiluminescence-based H2O2 measurements on a commercially available flow injection analysis (FIA) system with sampling of the analyte using microdialysis probes (MDPs), typically used for measurements in tissue. This allows minimally invasive, quantitative measurements of extracellular H2O2 concentration and dynamics utilizing the chemiluminescent reaction of H2O2 with acridinium ester. By coupling MDPs to the FIA system, measurements are no longer limited to filtered, liquid samples with low viscosity, as sampling via a MDP is based on a dynamic exchange through a permeable membrane with a specific cut-off. This allows continuous monitoring of dynamic changes in H2O2 concentrations, alleviates potential pH effects on the measurements, and allows for flexible application in different media and systems. We give a detailed description of the novel experimental setup and its measuring characteristics along with examples of application in different media and organisms to highlight its broad applicability, but also to discuss current limitations and challenges. The combined FIA-MDP approach for H2O2 quantification was used in different biological systems ranging from marine biology, using the model organism Exaiptasia pallida (light stress induced H2O2 release up to ~ 2.7 µM), over biomedical applications quantifying enzyme dynamics (glucose oxidase in a glucose solution producing up to ~ 60 µM H2O2 and the subsequent addition of catalase to monitor the H2O2 degradation process) and the ability of bacteria to modify their direct environment by regulating H2O2 concentrations in their surrounding media. This was shown by the bacteria Pseudomonas aeruginosa degrading ~ 18 µM background H2O2 in LB-broth. We also discuss advantages and current limitations of the FIA-MDP system, including a discussion of potential cross-sensitivity and interfering chemical species.

Original languageEnglish
JournalFree Radical Biology and Medicine
Volume128
Pages (from-to)111-123
Number of pages13
ISSN0891-5849
DOIs
Publication statusPublished - 20 Nov 2018

Bibliographical note

Copyright © 2018 Elsevier Inc. All rights reserved.

    Research areas

  • Chemiluminescence, Extracellular measurement, FIA, Flow injection analysis, HO, Hydrogen peroxide, Microdialysis probe, Optical sensor, ROS

See relations at Aarhus University Citationformats

ID: 130781028