T1 - Effect of Post-Dilatation Following Primary PCI With Everolimus-Eluting Bioresorbable Scaffold Versus Everolimus-Eluting Metallic Stent Implantation
T2 - An Angiographic and Optical Coherence Tomography TROFI II Substudy
AU - Yamaji, Kyohei
AU - Brugaletta, Salvatore
AU - Sabaté, Manel
AU - Iñiguez, Andrés
AU - Jensen, Lisette Okkels
AU - Cequier, Angel
AU - Hofma, Sjoerd H
AU - Christiansen, Evald Høj
AU - Suttorp, Maarten
AU - van Es, Gerrit Anne
AU - Sotomi, Yohei
AU - Onuma, Yoshinobu
AU - Serruys, Patrick W
AU - Windecker, Stephan
AU - Räber, Lorenz
PY - 2017/9/25
Y1 - 2017/9/25
N2 - Objectives This study sought to investigate the effect of post-dilatation on angiographic and intracoronary imaging parameters in the setting of primary percutaneous coronary intervention comparing the everolimus-eluting bioresorbable scaffold (BRS) with the everolimus-eluting metallic stent (EES). Background Routine post-dilatation of BRS has been suggested to improve post-procedural angiographic and subsequent device-related clinical outcomes. Methods In the ABSORB STEMI TROFI II trial, 191 patients with ST-segment elevation myocardial infarction were randomly assigned to treatment with BRS (n = 95) or EES (n = 96). Minimal lumen area and healing score as assessed by optical coherence tomography at 6 months were compared between BRS- and EES-treated patients stratified according to post-dilatation status. Results Primary percutaneous coronary intervention with post-dilatation was performed in 48 (50.5%) BRS- and 25 (25.5%) EES-treated lesions. There were no differences in baseline characteristics and post-procedural minimal lumen diameter between groups. In the BRS group, lesions with post-dilatation were associated with a trend toward a smaller minimal lumen area at 6 months (5.07 ± 1.68 mm
2 vs. 5.72 ± 1.77 mm
2; p = 0.09) and significantly larger angiographic late lumen loss (0.28 ± 0.34 mm vs. 0.12 ± 0.25 mm; p = 0.02), whereas no difference was observed in the EES arm (5.46 ± 2.18 mm
2 vs. 5.55 ± 1.77 mm
2; p = 0.85). The neointimal healing score was low and comparable between groups with and without post-dilation (BRS: 1.55 ± 2.61 vs. 1.92 ± 2.17; p = 0.48; EES: 2.50 ± 3.33 vs. 2.90 ± 4.80; p = 0.72). Conclusions In the setting of selected patients with ST-segment elevation myocardial infarction undergoing primary percutaneous coronary intervention with BRS or EES, post-dilatation did not translate into larger lumen area or improved arterial healing at follow-up.
AB - Objectives This study sought to investigate the effect of post-dilatation on angiographic and intracoronary imaging parameters in the setting of primary percutaneous coronary intervention comparing the everolimus-eluting bioresorbable scaffold (BRS) with the everolimus-eluting metallic stent (EES). Background Routine post-dilatation of BRS has been suggested to improve post-procedural angiographic and subsequent device-related clinical outcomes. Methods In the ABSORB STEMI TROFI II trial, 191 patients with ST-segment elevation myocardial infarction were randomly assigned to treatment with BRS (n = 95) or EES (n = 96). Minimal lumen area and healing score as assessed by optical coherence tomography at 6 months were compared between BRS- and EES-treated patients stratified according to post-dilatation status. Results Primary percutaneous coronary intervention with post-dilatation was performed in 48 (50.5%) BRS- and 25 (25.5%) EES-treated lesions. There were no differences in baseline characteristics and post-procedural minimal lumen diameter between groups. In the BRS group, lesions with post-dilatation were associated with a trend toward a smaller minimal lumen area at 6 months (5.07 ± 1.68 mm
2 vs. 5.72 ± 1.77 mm
2; p = 0.09) and significantly larger angiographic late lumen loss (0.28 ± 0.34 mm vs. 0.12 ± 0.25 mm; p = 0.02), whereas no difference was observed in the EES arm (5.46 ± 2.18 mm
2 vs. 5.55 ± 1.77 mm
2; p = 0.85). The neointimal healing score was low and comparable between groups with and without post-dilation (BRS: 1.55 ± 2.61 vs. 1.92 ± 2.17; p = 0.48; EES: 2.50 ± 3.33 vs. 2.90 ± 4.80; p = 0.72). Conclusions In the setting of selected patients with ST-segment elevation myocardial infarction undergoing primary percutaneous coronary intervention with BRS or EES, post-dilatation did not translate into larger lumen area or improved arterial healing at follow-up.
KW - Journal Article
UR - https://www.scopus.com/pages/publications/85029587831
U2 - 10.1016/j.jcin.2017.07.035
DO - 10.1016/j.jcin.2017.07.035
M3 - Journal article
C2 - 28935079
SN - 1936-8798
VL - 10
SP - 1867
EP - 1877
JO - JACC. Cardiovascular interventions
JF - JACC. Cardiovascular interventions
IS - 18
ER -