Dose response of three-dimensional silicone-based radiochromic dosimeters for photon irradiation in the presence of a magnetic field

Research output: Contribution to journal/Conference contribution in journal/Contribution to newspaperComment/debate/letter to the editorResearchpeer-review

The magnetic field in magnetic resonance imaging guided radiotherapy (MRgRT) delivery systems influences charged-particle trajectories and hence the three-dimensional (3D) radiation dose distributions. This study investigated the dose-response as well as dose-rate and fractionation dependencies of silicone-based 3D radiochromic dosimeters for photon irradiation in a magnetic field using a 0.35 T MRgRT system. We found a linear dose response up to 22.6 Gy and no significant dose-rate dependency as a function of depth. A difference in optical response was observed for dosimeters irradiated in a single compared to multiple fractions. The dosimeter showed clinical potential for verification of MRgRT delivery.
Original languageEnglish
JournalPhysics and Imaging in Radiation Oncology
Volume16
Pages (from-to)81-84
Number of pages4
DOIs
Publication statusPublished - Oct 2020

    Research areas

  • 3D radiation dosimetryMagnetic resonance imaging guided radiotherapy

See relations at Aarhus University Citationformats

ID: 199437431