Creating a clinical platform for carbon-13 studies using the sodium-23 and proton resonances

Research output: Contribution to journal/Conference contribution in journal/Contribution to newspaperJournal articleResearchpeer-review

DOI

  • James T Grist, University of Cambridge, University of Birmingham
  • ,
  • Esben S S Hansen
  • Juan D Sánchez-Heredia, Technical University of Denmark
  • ,
  • Mary A McLean, University of Cambridge
  • ,
  • Rasmus Tougaard
  • Frank Riemer, University of Cambridge
  • ,
  • Rolf F Schulte, GE healthcare, Freisinger Landstraße 50, 85748, Munich, Germany.
  • ,
  • Joshua D Kaggie, University of Cambridge
  • ,
  • Jan Henrik Ardenkjaer-Larsen, GE Healthcare, Chicago, USA, Technical University of Denmark
  • ,
  • Christoffer Laustsen
  • Ferdia A Gallagher, University of Cambridge

PURPOSE: Calibration of hyperpolarized 13 C-MRI is limited by the low signal from endogenous carbon-containing molecules and consequently requires 13 C-enriched external phantoms. This study investigated the feasibility of using either 23 Na-MRI or 1 H-MRI to calibrate the 13 C excitation.

METHODS: Commercial 13 C-coils were used to estimate the transmit gain and center frequency for 13 C and 23 Na resonances. Simulations of the transmit B1 profile of a Helmholtz loop were performed. Noise correlation was measured for both nuclei. A retrospective analysis of human data assessing the use of the 1 H resonance to predict [1-13 C]pyruvate center frequency was also performed. In vivo experiments were undertaken in the lower limbs of 6 pigs following injection of hyperpolarized 13 C-pyruvate.

RESULTS: The difference in center frequencies and transmit gain between tissue 23 Na and [1-13 C]pyruvate was reproducible, with a mean scale factor of 1.05179 ± 0.00001 and 10.4 ± 0.2 dB, respectively. Utilizing the 1 H water peak, it was possible to retrospectively predict the 13 C-pyruvate center frequency with a standard deviation of only 11 Hz sufficient for spectral-spatial excitation-based studies.

CONCLUSION: We demonstrate the feasibility of using the 23 Na and 1 H resonances to calibrate the 13 C transmit B1 using commercially available 13 C-coils. The method provides a simple approach for in vivo calibration and could improve clinical workflow.