Corrigendum to: highResNPS.com: An Online Crowd-Sourced HR-MS Database for Suspect and Non-targeted Screening of New Psychoactive Substances (Journal of Analytical Toxicology (2019) (1-8) DOI: 10.1093/jat/bkz030)

Research output: Contribution to journal/Conference contribution in journal/Contribution to newspaperComment/debateResearchpeer-review

DOI

  • Marie Mardal, Københavns Universitet
  • ,
  • Mette Findal Andreasen
  • Christian Brinch Mollerup, Københavns Universitet
  • ,
  • Peter Stockham, Forensic Science SA
  • ,
  • Rasmus Telving
  • Nikolaos S. Thomaidis, University of Athens
  • ,
  • Konstantina S. Diamanti, University of Athens
  • ,
  • Kristian Linnet, Københavns Universitet
  • ,
  • Petur Weihe Dalsgaard, Københavns Universitet

In the Paper: “HighResNPS.com: An Online Crowd-Sourced HR-MS Database for Suspect and Non-targeted Screening of New Psychoactive Substances”, the listed ‘Acquisition control’, ‘Collision energies’, and ‘Scan range(s)’ settings for the Agilent MassHunter screening in table II are incorrect. The correct acquisition control is: “DIA (AllIonsMS. Full-scan at four uccessive collision energies acquired over the 12 min chromatographic run. Scan rate at 5 spectra/sec, 0.8sec cycle time)”, Collision energies: “0V, 10V, 20V, and 40V”, and Scan range(s): “40-1,050”. The corrected Table II is reproduced here. The authors regret the errors.

Original languageEnglish
JournalJournal of Analytical Toxicology
Volume43
Issue6
Pages (from-to)E7-E8
Number of pages2
ISSN0146-4760
DOIs
Publication statusPublished - 24 Jul 2019

See relations at Aarhus University Citationformats

ID: 172984896