T1 - Clinical Evaluation of a New Spectral-Domain Optical Coherence Tomography-Based Biometer
AU - Alió, Jorge L.
AU - José-Martínez, Marina
AU - Martínez-Abad, Antonio
AU - Rodríguez, Alejandra E.
AU - Versaci, Francesco
AU - Hjortdal, Jesper
AU - Murta, Joaquim Neto
AU - Plaza-Puche, Ana B.
AU - Cantó-Cerdán, Mario
AU - Piñero, David P.
PY - 2024/3
Y1 - 2024/3
N2 - The VEMoS-AXL system is a new optical biometer based on spectral domain optical coherence tomography (SD-OCT) that has been tested in terms of intrasession repeatability and compared with a swept-source optical coherence tomography biometer (SS-OCT), which is recognized as the gold standard for the performance of an agreement analysis. A biometric analysis was performed three consecutive times in 120 healthy eyes of 120 patients aged between 18 and 40 years with the SD-OCT system, and afterwards, a single measurement was obtained with the SS-OCT system. Within-subject standard deviations were 0.004 mm, 4.394 µm, and 0.017 mm for axial length (AL), central corneal thickness (CCT), and anterior chamber depth (ACD) measures obtained with the SD-OCT biometer, respectively. The agreement between devices was good for AL (limits of agreement, LoA: −0.04 to 0.03 mm) and CCT (LoA: −4.36 to 14.38 µm), whereas differences between devices were clinically relevant for ACD (LoA: 0.03 to 0.21 mm). In conclusion, the VEMoS-AXL system provides consistent measures of anatomical parameters, being most of them interchangeable with those provided by the SS-OCT-based gold standard.
AB - The VEMoS-AXL system is a new optical biometer based on spectral domain optical coherence tomography (SD-OCT) that has been tested in terms of intrasession repeatability and compared with a swept-source optical coherence tomography biometer (SS-OCT), which is recognized as the gold standard for the performance of an agreement analysis. A biometric analysis was performed three consecutive times in 120 healthy eyes of 120 patients aged between 18 and 40 years with the SD-OCT system, and afterwards, a single measurement was obtained with the SS-OCT system. Within-subject standard deviations were 0.004 mm, 4.394 µm, and 0.017 mm for axial length (AL), central corneal thickness (CCT), and anterior chamber depth (ACD) measures obtained with the SD-OCT biometer, respectively. The agreement between devices was good for AL (limits of agreement, LoA: −0.04 to 0.03 mm) and CCT (LoA: −4.36 to 14.38 µm), whereas differences between devices were clinically relevant for ACD (LoA: 0.03 to 0.21 mm). In conclusion, the VEMoS-AXL system provides consistent measures of anatomical parameters, being most of them interchangeable with those provided by the SS-OCT-based gold standard.
KW - anterior chamber depth
KW - axial length
KW - central corneal thickness
KW - optical biometry
KW - spectral-domain optical coherence tomography
KW - swept-source optical coherence tomography