T1 - Assessing lumbar paraspinal muscle cross-sectional area and fat composition with T1 versus T2-weighted magnetic resonance imaging
T2 - Reliability and concurrent validity
AU - Cooley, J. R.
AU - Hebert, J. J.
AU - De Zoete, A.
AU - Jensen, T. S.
AU - Algra, P. R.
AU - Kjær, Per
AU - Walker, B. F.
N1 - Publisher Copyright:
© 2021 Cooley et al. This is an open access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited.
PY - 2021/2
Y1 - 2021/2
N2 - Purpose Studies using magnetic resonance imaging to assess lumbar multifidus cross-sectional area frequently utilize T1 or T2-weighted sequences, but seldom provide the rationale for their sequence choice. However, technical considerations between their acquisition protocols could impact on the ability to assess lumbar multifidus anatomy or its fat/muscle distinction. Our objectives were to examine the concurrent validity of lumbar multifidus morphology measures of T2 compared to T1-weighted sequences, and to assess the reliability of repeated lumbar multifidus measures. Methods The lumbar multifidus total cross-sectional area of 45 patients was measured bilaterally at L4 and L5, with histogram analysis determining the muscle/fat threshold values per muscle. Images were later re-randomized and re-assessed for intra-rater reliability. Matched images were visually rated for consistency of outlining between both image sequences. Bland-Altman bias, limits of agreement, and plots were calculated for differences in total cross-sectional area and percentage fat between and within sequences, and intra-rater reliability analysed. Results T1-weighted total cross-sectional area measures were systematically larger than T2 (0.2 cm2), with limits of agreement <±10% at both spinal levels. For percentage fat, no systematic bias occurred, but limits of agreement approached ±15%. Visually, muscle outlining was consistent between sequences, with substantial mismatches occurring in <5% of cases. Intra-rater reliability was excellent (ICC: 0.981-0.998); with bias and limits of agreement less than 1% and ±5%, respectively. Conclusion Total cross-sectional area measures and outlining of muscle boundaries were consistent between sequences, and intra-rater reliability for total cross-sectional area and percentage fat was high indicating that either MRI sequence could be used interchangeably for this purpose. However, further studies comparing the accuracy of various methods for distinguishing fat from muscle are recommended.
AB - Purpose Studies using magnetic resonance imaging to assess lumbar multifidus cross-sectional area frequently utilize T1 or T2-weighted sequences, but seldom provide the rationale for their sequence choice. However, technical considerations between their acquisition protocols could impact on the ability to assess lumbar multifidus anatomy or its fat/muscle distinction. Our objectives were to examine the concurrent validity of lumbar multifidus morphology measures of T2 compared to T1-weighted sequences, and to assess the reliability of repeated lumbar multifidus measures. Methods The lumbar multifidus total cross-sectional area of 45 patients was measured bilaterally at L4 and L5, with histogram analysis determining the muscle/fat threshold values per muscle. Images were later re-randomized and re-assessed for intra-rater reliability. Matched images were visually rated for consistency of outlining between both image sequences. Bland-Altman bias, limits of agreement, and plots were calculated for differences in total cross-sectional area and percentage fat between and within sequences, and intra-rater reliability analysed. Results T1-weighted total cross-sectional area measures were systematically larger than T2 (0.2 cm2), with limits of agreement <±10% at both spinal levels. For percentage fat, no systematic bias occurred, but limits of agreement approached ±15%. Visually, muscle outlining was consistent between sequences, with substantial mismatches occurring in <5% of cases. Intra-rater reliability was excellent (ICC: 0.981-0.998); with bias and limits of agreement less than 1% and ±5%, respectively. Conclusion Total cross-sectional area measures and outlining of muscle boundaries were consistent between sequences, and intra-rater reliability for total cross-sectional area and percentage fat was high indicating that either MRI sequence could be used interchangeably for this purpose. However, further studies comparing the accuracy of various methods for distinguishing fat from muscle are recommended.
UR - http://www.scopus.com/inward/record.url?scp=85100502581&partnerID=8YFLogxK
U2 - 10.1371/journal.pone.0244633
DO - 10.1371/journal.pone.0244633
M3 - Journal article
C2 - 33544707
AN - SCOPUS:85100502581
SN - 1932-6203
VL - 16
JO - PLOS ONE
JF - PLOS ONE
IS - 2
M1 - e0244633
ER -