PURPOSE: Left ventricular (LV) lead complications in cardiac resynchronization therapy are challenging and poorly reported. We aimed to establish prevalence, causes and outcomes of LV lead complications requiring re-intervention.
METHODS: We analysed the rate of complications in 2551 consecutive patients who received a transvenous de novo LV lead as part of a cardiac resynchronization therapy device between 2000 and 2018. LV lead complications requiring re-intervention were identified; those due to infection were excluded. Patient, procedural and device characteristics, and outcomes were examined for non-infective LV lead complications requiring re-intervention.
RESULTS: During a median of 4.7 years, 142 (5.6%) patients required re-intervention for non-infective LV lead complications with a decrease from 10.7% between 2000 and 2004, 8.7% between 2005 and 2009, 3.2% between 2010 and 2014 to 3.2% after 2014. The most common complications were LV lead displacement (50%), high pacing threshold (28%) and phrenic nerve stimulation (15%). Of the complications, 79 (56%) occurred within 90 days post-implant and 63 (44%) later. At the end of the study period, 132/142 patients (93%) had a functional LV lead. Lead re-intervention was associated with higher risk of complications (20%), but no increase in mortality (P = 0.19). Quadripolar leads had longer longevity and lower risk of complications compared with unipolar and bipolar LV leads.
CONCLUSIONS: A small but significant proportion of patients required LV lead re-intervention for complications following de novo implant. Lead displacement accounted for half of the re-interventions. Re-intervention was associated with a higher complication rate, but 92% of these patients had functional LV leads at the end of follow-up.