Mads Brix Kronborg

An unusually late dislodged atrial lead catching externalized ICD-lead conductor. A rare cause of simultaneous atrial and ICD lead over-sensing

Research output: Contribution to journal/Conference contribution in journal/Contribution to newspaperJournal articleResearchpeer-review

DOI

This case report demonstrates a late atrial lead dislodgement, ICD lead externalization and lead interaction due to the atrial lead screw-helix catching the externalized loop of the ICD lead, leading to simultaneous dual lead over-sensing in the post-shock phase.

Original languageEnglish
JournalJournal of arrhythmia
Volume35
Issue2
Pages (from-to)307-310
Number of pages4
ISSN1880-4276
DOIs
Publication statusPublished - Apr 2019

    Research areas

  • FAILURE, ICD over-sensing, inappropriate ICD therapy, lead dislodgement lead externalization, lead-complication