Christoffer Tobias Witt

Time until diagnosis of clinical events with different remote monitoring systems in Implantable Cardioverter-Defibrillator patients

Research output: Contribution to journal/Conference contribution in journal/Contribution to newspaperJournal articleResearchpeer-review

BACKGROUND: Remote monitoring (RM) is an established technology integrated into routine follow-up of patients with implantable cardioverter-defibrillator (ICD). Current RM systems differ according to transmission frequency and alert definition.

OBJECTIVE: We aimed to compare time difference between detection and acknowledgement of clinically relevant events between four RM systems.

METHODS: We analyzed time delay between detection of ventricular arrhythmic and technical events by the ICD and acknowledgement by hospital staff in 1.802 consecutive patients followed with RM during September 2014 - August 2016. Devices from Biotronik (BIO, n=374), Boston Scientific (BSC, n=196), Medtronic (MDT, n=468) and St Jude Medical (SJM, n=764) were included. We identified all events from RM webpages and their acknowledgement with RM or at in-clinic follow-up. Events occurring during weekends were excluded.

RESULTS: We included 3.472 events. Proportion of events acknowledged within 24 hours was 72%, 23%, 18% and 65% with BIO, BSC, MDT and SJM, respectively, with median times of 13, 222, 163 and 18 hours from detection to acknowledgement (p<0.001 for both comparisons between manufacturers). Including only events transmitted as alerts by RM, 72%, 68%, 61% and 65% for BIO, BSC, MDT and SJM, respectively were acknowledged within 24 hours. Variation in time to acknowledgement of ventricular tachyarrhythmia episodes not treated with shock therapy was the primary cause for the difference between manufacturers.

CONCLUSION: Significant and clinically relevant differences in time delay from event detection to acknowledgement exist between RM systems. Varying definitions of which events RM transmits as alerts are important for the differences observed.

Original languageEnglish
JournalHeart Rhythm
Volume15
Issue11
Pages (from-to)1648-1654
Number of pages7
ISSN1547-5271
DOIs
Publication statusPublished - Nov 2018