Abstract
We report here on the radiocarbon performance on the AARAMS HVE 1MV Tandetron. C-14 analysis is carried out in charge state 2+. We have avoided Li interference by appropriate settings of the high-energy electrostatic analyzer and the 30 degrees second high-energy magnet. The C-14 machine background is determined using unprocessed graphite, which yielded 58,650 +/- 2032 C-14 yr determined as the average and standard deviation of four measurements. International standards, which are used to monitor the long-term performance of the C-14 measurements, agree with the reported consensus values.
Originalsprog | Engelsk |
---|---|
Tidsskrift | Radiocarbon: An International Journal of Cosmogenic Isotope Research |
Vol/bind | 59 |
Nummer | 3 |
Sider (fra-til) | 905-913 |
Antal sider | 9 |
ISSN | 0033-8222 |
DOI | |
Status | Udgivet - jun. 2017 |