Monitoring of Free Flaps with Combined Tissue Spectrophotometry and Laser Doppler Flowmetry in an Animal Experimental Model

Publikation: Bidrag til tidsskrift/Konferencebidrag i tidsskrift /Bidrag til avisTidsskriftartikelForskningpeer review

DOI

  • Mette Marie Berggren-Olsen, Univ Southern Denmark, University of Southern Denmark, Inst Reg Hlth Res
  • ,
  • Andreas Rauff-Mortensen
  • ,
  • Rene Holst, Univ Oslo, University of Oslo, Oslo Ctr Biostat & Epidemiol
  • ,
  • Kim Christian Houlind, Univ Southern Denmark, University of Southern Denmark, Inst Reg Hlth Res
  • ,
  • Hanne Birke-Sorensen

Background When mobilizing free flaps, postoperative monitoring of perfusion is crucial to detect ischemia. Continuous monitoring may be feasible by applying a combination of tissue spectrophotometry and laser Doppler flowmetry (oxygen-2-see [O2C]).

Material and Methods On 10 pigs, two symmetrical myocutaneous flaps were mobilized on each side of the abdomen based on the deep inferior epigastric vessels. Flaps were randomized to clamp either the artery or the vein and measurements using O2C were performed before, during, and after the intervention yielding information on blood flow, saturation (sat), and relative tissue hemoglobin (rHgb) concentration.

Results Baseline values were similar in all groups. Introduction of ischemia caused a rapid decline in arterial ischemic flaps which all reached threshold levels in 3 minutes, whereas that was only the case for three of six venous ischemic flaps. Venous clamping resulted in a decline in sat, while the response to arterial clamping was an initial decline followed by an increase in sat. In all arterial ischemic flaps, rHgb concentration either decreased or remained at baseline levels but increased in all venous ischemic flaps. The median time to a 30% rise was 1 minute at an 8-mm depth. The rate of decreasing flow along with the rHgb measurements made it possible to distinguish the arterial ischemia (AI) from the venous ischemia (VI) within the first few minutes.

Conclusion In this animal experimental model, O2C measurements of blood flow reliably detected ischemia. By adding information about rHgb, it was possible to distinguish between AI and VI.

OriginalsprogEngelsk
TidsskriftJournal of Reconstructive Microsurgery
Vol/bind33
Nummer8
Sider (fra-til)579-586
Antal sider8
ISSN0743-684X
DOI
StatusUdgivet - okt. 2017

Se relationer på Aarhus Universitet Citationsformater

ID: 121736445