Comparison of Patent Foramen Ovale Sizing by Transesophageal Echocardiography and Balloon Sizing in Patients Undergoing Percutaneous Closure

Birgitte Carbuhn Larsen, Martin Christian Harmsen, Henrik Vase, Jens Erik Nielsen-Kudsk, Asger Andersen

Publikation: Bidrag til tidsskrift/Konferencebidrag i tidsskrift /Bidrag til avisTidsskriftartikelForskningpeer review

Abstract

Background: A patent foramen ovale (PFO) has a complex anatomy, and evaluating the size before closure may be challenging. We aimed to investigate the correlation between preprocedural transesophageal echocardiography (TEE) and balloon sizing of PFO in patients undergoing percutaneous PFO closure. Methods: A retrospective single-center study with analysis of 100 patients who, due to paradox thromboembolism in the left circulation, underwent percutaneous PFO closure. The PFO sizing was compared to measures attained by TEE and balloon sizing using linear regression analysis. Results: PFO size measured by TEE occurred smaller than balloon sizing (2.19 mm [95% CI: 1.91 to 2.46] vs. 8.51 mm [95% CI: 8.02 to 9.00], p < 0.001). Additionally, neither the PFO channel length nor the atrial septal mobility measured by TEE correlated to the PFO size attained by balloon sizing, respectively (slope −0.018 [95% CI: −0.117 to 0.081], R = 0.036, p = 0.719) and (slope 0.049 [95% CI: ?0.043 to 0.141], R = 0.105, p = 0.297). Statistically significant difference in regression analysis but poor correlation was found between both TEE attained PFO and shunt size when compared to balloon sizing. Diverting patients according to the size of the PFO shunt was not statistically significant between PFO of moderate size compared, respectively, to a large and small PFO size. However, a difference was observed between a small and large PFO shunt size. Conclusions: PFO defect and shunt size measured by TEE showed a poor correlation with balloon sizing. Neither PFO channel length nor septal mobility were correlated to the PFO size measured by balloon sizing.

OriginalsprogEngelsk
Artikelnummer100375
TidsskriftStructural heart : the journal of the Heart Team
Vol/bind9
Nummer1
Sider (fra-til)100375
ISSN2474-8706
DOI
StatusUdgivet - jan. 2025

Fingeraftryk

Dyk ned i forskningsemnerne om 'Comparison of Patent Foramen Ovale Sizing by Transesophageal Echocardiography and Balloon Sizing in Patients Undergoing Percutaneous Closure'. Sammen danner de et unikt fingeraftryk.

Citationsformater