TY - JOUR
T1 - Clinical outcomes of the ACURATE neo2 transcatheter heart valve
T2 - a prospective, multicentre, observational, post-market surveillance study
AU - Kim, Won-Keun
AU - Tamburino, Corrado
AU - Möllmann, Helge
AU - Montorfano, Matteo
AU - Ellert-Gregersen, Julia
AU - Rudolph, Tanja K
AU - Van Mieghem, Nicolas M
AU - Hilker, Michael
AU - Amat-Santos, Ignacio J
AU - Terkelsen, Christian Juhl
AU - Petronio, Anna Sonia
AU - Stella, Pieter R
AU - Götberg, Matthias
AU - Rück, Andreas
AU - Kasel, A Markus
AU - Trillo, Ramiro
AU - Appleby, Clare
AU - Barbanti, Marco
AU - Blanke, Philipp
AU - Modolo, Rodrigo
AU - Allocco, Dominic J
AU - Sondergaard, Lars
PY - 2023/5
Y1 - 2023/5
N2 - Background: The next-generation ACURATE neo2 transcatheter aortic valve was designed for simplified implantation and to mitigate the risk of paravalvular leak (PVL) compared to the earlier device. Aims: We sought to collect clinical outcomes and device performance data, including echocardiography and 4-dimensional computed tomography (4D-CT) data, with the ACURATE neo2 transcatheter heart valve in patients with severe aortic stenosis (AS). Methods: The ACURATE neo2 Post-Market Clinical Follow-up (PMCF) Study is a single-arm, multicentre study of patients with severe AS treated in routine clinical practice. The primary safety endpoint was all-cause mortality at 30 days. The primary imaging endpoint was hypoattenuated leaflet thickening (HALT), measured by core laboratory-adjudicated 4D-CT at 30 days. Secondary endpoints included Valve Academic Research Consortium safety endpoints, procedural success, and evaluation of valve performance via core laboratory-adjudicated echocardiography. Results: The study enrolled 250 patients at 18 European centres (mean age: 80.8 years; 63.6% female; mean Society of Thoracic Surgeons score: 2.9±2.0%); 246 patients (98.4%) were successfully treated with the ACURATE neo2. The 30-day rates for mortality and disabling stroke were 0.8% and 0%, respectively. The new permanent pacemaker implantation rate was 6.5%. HALT >50% was present in 9.3% of patients at 30 days. Valve haemodynamics improved from baseline to 30 days (mean aortic valve gradient: from 47.6±14.5 mmHg to 8.6±3.9 mmHg; mean aortic valve area: from 0.7±0.2 cm2 to 1.6±0.4 cm2). At 30 days, PVL was evaluated as none/trace in 79.2% of patients, mild in 18.9%, moderate in 1.9%, and severe in 0%. Conclusions: The study results support the safety and efficacy of transcatheter aortic valve implantation with the ACURATE neo2 in patients in routine clinical practice.
AB - Background: The next-generation ACURATE neo2 transcatheter aortic valve was designed for simplified implantation and to mitigate the risk of paravalvular leak (PVL) compared to the earlier device. Aims: We sought to collect clinical outcomes and device performance data, including echocardiography and 4-dimensional computed tomography (4D-CT) data, with the ACURATE neo2 transcatheter heart valve in patients with severe aortic stenosis (AS). Methods: The ACURATE neo2 Post-Market Clinical Follow-up (PMCF) Study is a single-arm, multicentre study of patients with severe AS treated in routine clinical practice. The primary safety endpoint was all-cause mortality at 30 days. The primary imaging endpoint was hypoattenuated leaflet thickening (HALT), measured by core laboratory-adjudicated 4D-CT at 30 days. Secondary endpoints included Valve Academic Research Consortium safety endpoints, procedural success, and evaluation of valve performance via core laboratory-adjudicated echocardiography. Results: The study enrolled 250 patients at 18 European centres (mean age: 80.8 years; 63.6% female; mean Society of Thoracic Surgeons score: 2.9±2.0%); 246 patients (98.4%) were successfully treated with the ACURATE neo2. The 30-day rates for mortality and disabling stroke were 0.8% and 0%, respectively. The new permanent pacemaker implantation rate was 6.5%. HALT >50% was present in 9.3% of patients at 30 days. Valve haemodynamics improved from baseline to 30 days (mean aortic valve gradient: from 47.6±14.5 mmHg to 8.6±3.9 mmHg; mean aortic valve area: from 0.7±0.2 cm2 to 1.6±0.4 cm2). At 30 days, PVL was evaluated as none/trace in 79.2% of patients, mild in 18.9%, moderate in 1.9%, and severe in 0%. Conclusions: The study results support the safety and efficacy of transcatheter aortic valve implantation with the ACURATE neo2 in patients in routine clinical practice.
KW - MSCT
KW - TAVI
KW - aortic stenosis
KW - femoral
KW - transoesophageal echocardiogram
UR - http://www.scopus.com/inward/record.url?scp=85163439846&partnerID=8YFLogxK
U2 - 10.4244/EIJ-D-22-00914
DO - 10.4244/EIJ-D-22-00914
M3 - Journal article
C2 - 36440588
SN - 1774-024X
VL - 19
SP - 83
EP - 92
JO - EuroIntervention : journal of EuroPCR in collaboration with the Working Group on Interventional Cardiology of the European Society of Cardiology
JF - EuroIntervention : journal of EuroPCR in collaboration with the Working Group on Interventional Cardiology of the European Society of Cardiology
IS - 1
ER -