Abstract
Previous calculations of the γ dose given to Risø Calibration Quartz (RCQ) were performed using mass absorption and mass attenuation coefficients. In this paper, we update the γ dose given to RCQ using Geant4 simulations and provide a comparison with three other γ sources in Denmark. We also show experimental evidence that the luminescence response to a137Cs γ irradiation is 4% higher than that resulting from a60Co γ irradiation for the same dose, confirming the suggestions of Autzen et al. (2021) based on simulations.
Originalsprog | Engelsk |
---|---|
Artikelnummer | 106828 |
Tidsskrift | Radiation Measurements |
Vol/bind | 157 |
ISSN | 1350-4487 |
DOI | |
Status | Udgivet - sep. 2022 |