Publikation: Bidrag til tidsskrift/Konferencebidrag i tidsskrift /Bidrag til avis › Tidsskriftartikel › Forskning › peer review
OBJECTIVE: To assess operator-, examination-, and patient-related factors, affecting patient movement and re-exposure in cone beam computed tomography (CBCT) examination.
STUDY DESIGN: The sample consisted of 248 CBCT examinations in 190 patients video-recorded during examination. Three observers scored the videos; the patient moved or did not move. Operator-, examination-, and patient-related factors were evaluated separately (chi-square test) and by multivariate regression analyses (patient movement and re-exposure as separate outcomes).
RESULTS: The prevalence of movement was 21%. Cotton roll stabilizing patient's jaws, CBCT unit touching patient's hair, and patient's age 15 years or greater were related to movement. Age 15 years or greater had a significant impact on movement (P < .001; odds ratio [OR] 11.0). There were 16 re-exposures (6.4%). Age 15 years or greater, presence of a cotton roll, and field of view (FOV) were related to re-exposure. Use of a large FOV had significant impact on re-exposure (P = .04; OR 5.8).
CONCLUSIONS: Operator-, examination-, and patient-related factors may affect patient movement and re-exposure in CBCT examination.
Originalsprog | Engelsk |
---|---|
Tidsskrift | Oral Surgery, Oral Medicine, Oral Pathology and Oral Radiology |
Vol/bind | 119 |
Nummer | 5 |
Sider (fra-til) | 572-578 |
Antal sider | 7 |
ISSN | 1079-2104 |
DOI | |
Status | Udgivet - 3 feb. 2015 |
Se relationer på Aarhus Universitet Citationsformater
ID: 85620256