Effect of Software Version on the Accuracy of an Intraoral Scanning Device

Publikation: Bidrag til tidsskrift/Konferencebidrag i tidsskrift /Bidrag til avisTidsskriftartikelForskningpeer review

DOI

PURPOSE: To investigate the impact of software version on the accuracy of an intraoral scanning device.

MATERIALS AND METHODS: A master tooth was scanned with a high-precision optical scanner and then 10 times with a CEREC Omnicam scanner with software versions 4.4.0 and 4.4.4. Discrepancies were measured using quality control software.

RESULTS: Mean deviation for 4.4.0 was 36.2 ± 35 μm and for 4.4.4 was 20.7 ± 14.2 μm (P ≤ .001).

CONCLUSION: Software version has a significant impact on the accuracy of an intraoral scanner. It is important that researchers also publish the software version of scanners when publishing their findings.