RADIOCARBON ANALYSIS ON THE NEW AARAMS 1MV TANDETRON

Publikation: Bidrag til tidsskrift/Konferencebidrag i tidsskrift /Bidrag til avisTidsskriftartikelForskningpeer review

DOI

We report here on the radiocarbon performance on the AARAMS HVE 1MV Tandetron. C-14 analysis is carried out in charge state 2+. We have avoided Li interference by appropriate settings of the high-energy electrostatic analyzer and the 30 degrees second high-energy magnet. The C-14 machine background is determined using unprocessed graphite, which yielded 58,650 +/- 2032 C-14 yr determined as the average and standard deviation of four measurements. International standards, which are used to monitor the long-term performance of the C-14 measurements, agree with the reported consensus values.

OriginalsprogEngelsk
TidsskriftRadiocarbon: An International Journal of Cosmogenic Isotope Research
Vol/bind59
Nummer3
Sider (fra-til)905-913
Antal sider9
ISSN0033-8222
DOI
StatusUdgivet - jun. 2017

Se relationer på Aarhus Universitet Citationsformater

ID: 114687459